5 Unpacking, installation and commissioning

5.1

Place of installation, place of use


The balance is constructed in such a way that reliable weighing results can be
achieved under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and
precisely.

Therefore please observe the following at the place of installation:

-

Place the balance on a firm, level surface;

-

Avoid extreme heat as well as temperature fluctuation caused by installing

next to a radiator or in the direct sunlight;

-

Protect the balance against direct draughts due to open windows and doors;

-

Avoid jarring during weighing;

-

Protect the balance against high humidity, vapours and dust;

-

Do not expose the device to extreme dampness for longer periods of time. In-

admissible bedewing (condensation of air moisture on the device) can occur if
a cold device is taken into a significantly warmer environment. In this case,
please acclimatise the device for approx. 2 hours at room temperature after it
has been disconnected from the mains.

-

Avoid static charging of the material to be weighed, weighing container and

windshield.


Major display deviations (incorrect weighing results) are possible if electromagnetic
fields occur as well as due to static charging and instable power supply. It is then
necessary to change the location.