5.7.3 Non-verifiable Models

FOB 1.5K0.5, FOB 3K1, FOB 6K2, FOB 7.5K0.5L, FOB 15K1L, FOB 30K2L

Ö Turn balance off and hold

key, test procedure appears on the display, op-

erate the

-key 3x times during this test procedure.

Ö Now let go of both keys. In the display appears “CAL”. Press key

in dis-

play appears “Load”.

Ö Place the tare weight (nominal value, see “Techn. Specifications“ chapter 1) in

the centre of the weighing pan, then press the key

. “CAL” will appear on

the display for approx. 2 seconds, followed by the nominal value of the adjust-
ing weight.

Ö Remove the adjusting weight from the weighing plate to complete adjusting.

5.7.4 Check software settings for verification of a balance

(only models FOB 1K1M, FOB 2K2M, FOB 5K5M):

For the adjustment function, the balance must be switched over to a service
mode. This service mode is activated by actuating the adjustment switch
twice.

In the service mode all parameters of the balance can be modified. There is a
special setting named „TRAD“.

The service parameters may not be modified, as this could damage the bal-
ance settings.

Especially the setting „TRAD“ is very important for a verification. This setting
activates and deactivates the verifiable settings of the balance.

Prior to a verification the setting „TRAD“ must be revised.

Press the

key and keep it pressed until „A_OFF“ appears on the display.

Press

button repeatedly. The setting „Trad“ must not appear (see

graphic).