Procedure when adjusting:

Observe stable environmental conditions.

A warming up time (see chapter 1; Technical Data) is required for stabilization.

Ensure that there are no objects on the weighing plate.

Models FOB-NS:


Balance display

Operation

(example)

 Switch off the balance

 Press and hold

, press

3x at the same

time, release

- an internal value is displayed.


(example)



 Press

, "0" is displayed

 Put adjustment weight in the centre of the weighing

plate and press

, "CAL" is displayed, followed

by the value of the adjustment weight.

 Take away adjustment weight
 The balance changes into weighing mode.
 Adjustment has now been completed successfully.