3 data output rs 232 c – KERN & SOHN CDS User Manual

Page 18

Advertising
background image

7.3 Data output RS 232 C

Technical Data

8-bit ASCII Code
 1 start bit, 8 data bits, 1 stop bit, no parity bit
 Baud rate selectable from 2400, 4800, 9600 Baud (factory setting) and 19200 Baud.
 Sub-D plug 9-channel required
 For operation with interface faultless operation is only ensured with the correct KERN – interface cable

(max. 2m)

Pin allocation of the balance output socket (front view)

Sub-D jack 9-channel

5

1

9 6

Pin 2:

Transmit data


Pin 3:

Receive data


Pin 5:

Signal ground


Baud rate

The MODE key is used to set the Baud rate used for the transfer of measured values. In the following example
the baud rate is set to 4800 baud.

Setting Baud rate KERN CKE

Display

1. Switch on balance
2. Press and hold tare key
3. Tap ON/OFF key and let go of tare key
"Cal?" will appear on the display.
4.

Keep holding the switch over key „pcs ↔ g“ until "Printer" appears on the
display and confirm by pressing the "20x" key. 2.400 Bd will appear on the
display.

5.

Press the switchover key "pcs ↔ g" to select the Baud rate and confirm

by the “20x” key. To return to weighing mode, press the tare key.

PRINTER?

2400

Baud

4800

Baud

4800 Baud X

0.00 oz