9.2 Numeric entering of the reference weight

If you know the reference weight/piece you can enter this via number keys.

Enter reference weight via number keys

0

W

a

PRE-

TARE

1.833

0

M+

W

Confirm with

key

0

W

a

PRE-

TARE

1.833

Reference weight

0

M+

W

Now you can place the parts to be counted onto the weighing plate. All quantity pa-
rameters of your goods to be weighed are displayed:

9.3 Automatic reference optimization

If it was impossible to determine a reference due to instable goods to be weighed or
an insufficient reference weight, the [W ] display will appear in the reference weight
window during reference calculation.

Overlay W indicates:

Placed number of pieces insufficient for reference calculation

CXB models < 40 d

CXP models < 20 d

Placed reference weight insufficient for reference calculation

CXB models < 4/5 d

CXP models < 1/5 d


Add additional parts until the [W ] display disappears.
An audio signal indicates that reference optimization has been carried out.
At every reference optimisation, the reference weight is calculated anew. As the addi-
tional pieces increase the base for the calculation, the reference also becomes more
exact.