In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.

Fault

Possible cause

The balance is not switched on.
The mains supply connection has been

interrupted (mains cable not plugged
in/faulty).

Power supply interrupted.
Batteries are inserted incorrectly or empty

The displayed weight does not glow.

No batteries inserted.
Draught/air movement

Table/floor vibrations

The displayed weight is permanently
changing

Weighing plate has contact with other

objects.

Electromagnetic fields / static charging

(choose different location/switch off inter-
fering device if possible)

The display of the balance is not at zero
Adjustment is no longer correct.

The weighing result is obviously in-
correct

Great fluctuations in temperature.

Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.