KERN & SOHN EOB F User Manual

Page 15

Advertising
background image

EOS_F-BA-e-1310

15


In weighing mode press the UNIT key and keep it pressed until „CAL“ followed

by the weight value of the required adjustment weight in displayed. „LOAd“ is

displayed alternatively.

(Example)

Carefully place the required adjustment weight (see tab. 1 „Technical data“) in

the centre of the weighing plate

Press TARE key, adjustment is started.

(Example)

Wait until the stability display appears, after that the adjustment is finished. The

balance will automatically return to Weighing mode upon successful completion

of the adjustment and will display the weight of the adjustment weight.

(Example)

Take away adjustment weight.

An error message will be displayed in the event of an adjustment error or

incorrect adjustment weight. Remove the adjustment weight and repeat the

adjustment process.

Keep the adjustment close to the balance. Daily control of the weighing

exactness is recommended for quality-relevant applications.

Advertising