7 Functions of display unit

7.1 Prior to weighing: ZERO-POINT CHECK
Prior to each weighing process ensure that the system is unloaded and clear of
obstructions. The display unit provides automatic zero adjustment. This means, that
small deviations from the zero-point are corrected automatically. If the display unit
does not correct the zero-point automatically you will have to carry out a manual
correction with the help of the

TARE

key.

7.2 Gross weighing
After the load was lifted, the display shows the gross value of the weighed weight.

7.3 Net weighing: TARING BY PRESSING A BUTTON
The display unit offers an option that allows tare weights to be reset to zero by
pressing a button. In this way net weight changes can be traced. After taring, the
display unit will once again start at the smallest display step.

¾

Lift the load.

¾

Press

TARE

button

‰

The display unit shows zero.

‰

The lit up display “NET“, indicates that a tare weight is active.

¾

Loading and unloading the net load.

‰

The net value for the weighed weight is shown on the display.

‰

For unloading this will be a negative value.

¾

The system reverts to standard weighing mode when a zero adjustment is carried

out in an unloaded state.

7.4 Net weighing: MANUAL TARE INPUT
You may enter the tare weight at any time, that is, regardless whether the scales are
in a loaded or unloaded state. To achieve higher accuracy, a tare weight with a
higher resolution may be entered, regardless of the size of the weight and the display
steps of the indicator.

¾

Press

PRE-

TARE

button

‰

The most recently used tare value appears.

‰

The segment on the right is flashing.

¾

Press the

TARE

key for three seconds if the indicated tare value is to be used again


or