20

HDB-N-BA-e-1320

5.7.1

Data-Hold function

 Switch on the balance, keep the HOLD-button pressed until the current setting

„Hx“ (H1 – H6) appears.

 Press ON/OFF button repeatedly until the desired setting

„H1-H4“ is displayed.

 Confirm setting with the HOLD button.
 Suspend the material to be weighed
 The weight value is fixed and displayed depending on the setting (H1

– H4) (see

chap. 5.7), represented by the [] in the upper left.

5.7.2

Animal weighing function

This function is suitable for busy weighing procedures. The result is a mean value
formed by 16 weighing values which is found out within 3 seconds.

 Switch on the balance, keep the HOLD-button pressed until the current setting

„Hx“ (H1 – H6) appears.

 Press ON/OFF button repeatedly until the setting

„H5“ is displayed.

 Confirm setting with the HOLD button.
 Suspend the material to be weighed
 Press the HOLD-button, the display counts reverse from 3 -1.

The calculated mean value is indicated, symbolised by the [] top left.

 Before further measurements press first the ON/OFF/TARE-button.