Overview:

Function

Available
settings

Description

Low

High*

cap

d (low)

d (high)

1 t

2kg

1kg

500g 500g 200g 100g

3

t 10kg 5kg 2kg 2kg 1kg 500g

5

t 10kg 5kg 2kg 2kg 1kg 500g

10

t 20kg 10kg 5kg 5kg 2kg 1kg

F0 di
Modification of readability

AZn 0

0.5 d

AZn 1*

1 d

AZn 2

2 d

F1 az
Automatic zero point
correction
(zero tracking)

AZn 3

4 d

F2 bt

Not documented

F3 sp

Not documented

F4 ip

Internal A/D converter value

F5 ut

Not documented

Modifications may only be carried out by a

specialist with competent knowledge.

SLP 0

Standby mode switched off

SLP 1*

Standby mode after 5 minutes

SLP 2

Standby mode after 10 minutes

SLP 3

Standby mode after 20 minutes

F6 sl
Standby mode
see chap. 4.3.1

SLP 4

Standby mode after 30 minutes

F7 gv

Not documented

F8 ci
Password entry

In „P1- - - “ display increase the flashing number using

, select number using HOLD. Confirm entry by

ZERO button.

F9 CL

Adjustment, see chapter 7

* = default setting