Procedure:

Ö

Call-up menu setting

, see chap. 1.

Ö

Acknowledge with

.

Ensure that there are no objects on the weighing plate.

Ш

Ц

Wait for stability display „STABLE“, then press

. When

“LoAd 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform.

Ö

Wait for stability display „STABLE“, then press

. When

“LoAd 2“ is displayed, put the second adjustment weight (2/3
max) carefully in the centre of the weighing platform.

Ö

Wait for stability display „STABLE“, then press

. When

“LoAd 3“ is displayed, put the third adjustment weight (1/3
max) carefully in the centre of the weighing platform.

Ö

Wait for stability display „STABLE“, then press

.

Ö

After the adjustment the balance will carry out a self-test.

Remove adjusting weight during selftest, the appliance will
return into weighing mode automatically.

An adjusting error or incorrect adjusting weight will be indi-
cated by the error message; repeat adjustment procedure.