11 Troubleshooting


The scale should be switched off for a short time following an interruption in the pro-
gram sequence and disconnected from mains supply. It is then necessary to repeat
the weighing process from the beginning.

Help:

Interruption Possible

cause

Weight display is not illumi-
nated.

• The scale is not switched on.

• Batteries are incorrectly inserted or discharged

• No

batteries.

Weight display changes
continuously

• Draught/air

movement

• Table/floor

vibrations

• The weighing plate is in contact with foreign mat-

ters or is installed incorrectly.

• Electromagnetic fields/static charging (choose a

different location for the scale, switch off an inter-
fering device if possible)

The weighing result is obvi-
ously incorrect

• The scale display is not set to zero

• Incorrect

adjustment.

• Great fluctuations in temperature .

• Warm-up time was ignored.

• Electromagnetic fields/static charging (choose a

different location for the scale, switch off an inter-
fering device if possible)





Should other error messages occur, switch the scale off and then on again. If the er-
ror message remains, inform the manufacturer.