Instruction manual su, Sauter gmbh – KERN & SOHN SU 130 User Manual

Page 3

Advertising
background image

Sauter GmbH

Ziegelei 1
D-72336 Balingen
E-Mail: [email protected]

Tel: +49-[0]7433- 9933-199
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com

Instruction Manual

SU

SU-BA-e-1311

3

5.2 No matter, in

or in

state, the memorized data

can be browsed by pressing the

key 3-17. The

browsing state is marked by `R` on the display.
In `R` state, all the readings memorized, can be recalled

by pressing the

or

key.


5.3 To delete the memorized value in memory, the
browsing state has just to be entered and the reading to be

deleted has to be located by pressing the

or

key.

Then the

key 3-21 has to be pressed. If there appears

“Err0” on the display, this indicates that there is no more
reading to delete.

6. How to set the “Leq” Time


Leq is used to access the rms average noise level over a
preset period of time, often the staring point of a noise
assessment. To take the Leq measurement, the period of
time which has to be tested, must be selected. The longer
the period of measurement time, the more accurate the
Leq reading will be. To set the period of measurement

time, the

key 3-21 just has to be pressed for about 8

seconds until “Leq” appears on the display. Then the key
has to be released and the time has to be selected among
10s, 1 min, 5 min, 10 min, 15 min, 30 min, 1 hour, 8 hours
or 24 hours by pressing the Up (3-16) or Down (3-17) key.

To quit, any key except

or

key can be pressed.


7. How to set the alarm value


The alarm LED is on when the instant measurement value
is bigger than or equal to the alarm value set. The default
setting value at the factory is 85dB. It can be changed to
any value between 30 and 130dB by the following steps:

The

key 3-21 has to be pressed for about 5 seconds.

It has to be released after “AL” is shown on the display.

Then

key or

key have to be pressed to preset the

alarm value as desired.

To quit, any key except

or

key can be pressed.


8. How to set the time of Auto Power off

The default setting for auto power off at the factory is 5
minutes, which means that the device will auto power off 5
minutes from the time of the last key operation. This can
be changed by the user to any value between 1 to 9
minutes by the following steps:

The

key 3-21 has to be pressed for about 10 seconds

and it can be released after “AUTO” is shown on the

display. Then the

or

key has to be pressed to

preset the time as desired. To stop the function of auto
power off, time just has to be preset to “0”. So the device

will not be powered off automatically, it can only be shut
down manually in that case.

9. Calibration


The standard method to calibrate the instrument requires
an external calibrator (ASU-01) in addition to a small
screw-driver.

9.1 Calibration of the device :
a) The device has to be turned on.
b) It has to be preset in the “A” weighting mode
c) It also has to be preset the “SLOW” response mode
d) Then the microphone has to be placed into the cali-
brator. The calibrator has to be switched ON.
e) The CAL potentiometer of the device has to be
adjusted in the way that the device’s display matches the
output of the calibrator.
9.2 Calibration of the device by the inbuilt signal
The instrument has to be switched on to enter the

calibration state. Then the

key 3-18 has to be

pressed until the icon “CAL” is to be seen on the display.
The screw-driver has to be used to adjust the CAL
potentiometer of the device in the way that the device
shows 94dB.

10. Considerations

10.1 Wind blowing across the microphone increases the
noise measurement. The supplied foam cover has to be
used to cover the microphone when applicable.
10.2 The instrument may not be dismantled by
inexperienced staff. There are no parts inside which might
serve the user.

11. Battery replacement

11.1 When the battery voltage is less than approx. 5V, it is
necessary to replace batteries.

11.2 The 4 x 1.5 V batteries have to be installed correctly
into the case .

11.3 If the device is not going to be used for an extended
period, batteries have to be removed.

Note: If the AC adapter is used, batteries should be taken

out.

12. Takeover of data memory into software


Important: First start the continuous data logging and then
read out the stored data!

1. Connect the SU 130 to the PC (cable is optional
available with software ATC 01).
2. Put the SU 130 into the Memory Read Mode (by
pressing the READ-button).