Press 【

】 or 【

】 key to move cursor to【 System

Set】 .
Press 【 ENTER】 key to enter 【 System Set】 menu. Press

】 or 【

】 key to move cursor to the item to be set.

Press 【 ENTER】 key to directly change, or enter
corresponding changing interface.
Press 【 ESC】 key to return.
As for the items of 【 Auto Save】 , 【 Auto Print】 , 【 Auto
Delete】 , 【 Auto Trans.】 , 【 Key Sound】 and 【 Warn.
Sound】 , 【 ENTER】 key can be pressed to selected
【 On】 or 【 Off】 .
If 【 Auto Save】 is set to 【 On】 , the current group data can
be saved automatically after measurement is finished and
the average value is displayed.
If 【 Auto Print】 is set to 【 On】 , the current data can be
printed out after measurement is finished and the average
value is displayed.
If 【 Auto Delete】 is set to 【 On】 , gross error can be
deleted automatically when average times had been
achieved or measurement had been finished in advance
by pressing 【 AVE.】 key according to 3

бrule. If some data

had been cancelled, additional measurements should be
carried out to satisfy the times set.
If 【 Auto Trans.】 is set to 【 On】 , the current group data
can be output in text format via RS232 after measuring
finished and average value displayed.
If 【 Key Sound】 is set to 【 On】 , buzzer will send out a
short sound with each pressing.
If 【 Warn. Sound】 is set to【 On】 , buzzer will send out a
long sound in case of measured value exceeds the
tolerance limit, data deletion or other things.
If 【 Auto Down】 is set to 【 On】 , Power will turn off if
neither measurement nor any key operation is performed
within 5 minutes.

6.9.1 LCD brightness Set

Press 【

】 key to increase brightness.

Press 【

】 key to reduce brightness.

Press 【 ENTER】 key to finish change.
Press 【 ESC】 key to cancel change.
The brighter of brightness, the deeper the colour; The
darker of brightness, the lighter the colour;

6.9.2 Time and Date Set

When in this interface, current time and date will be
displayed on the screen, the format is “mm/dd/yy”.

Press 【

】 or 【

】 key to input the value, Press

】 or【

】 key to move cursor.

Press 【 ENTER】 key to finish the change, which current
time and date will be replaced by time and date set. Press
【 ESC】 key to cancel the change.

6.10 About software

When in the main display interface, press 【 MENU】 key to
enter the main menu.

Press 【

】 or 【

】 key to move cursor to 【 About

Software】 .
Press 【 ENTER】 key to enter 【 About Software】 .

Information about the tester and embedded software will
be displayed on this interface.
The software version and embedded software identification
are subjected to change due to the upgrading of software
without notifying in advance.


6.11 Calibration of the Tester

The tester and impact device must be calibrated with a
standard Leeb hardness test block (not included in
delivery) prior to the first use, or reusing after a long term
idle.
One time calibration is enough for each type of impact
device which equipped with a main unit; recalibration is not
necessary after the replacement of impact device later.

Press 【 ENTER】 key as well as

key at the same time to

enter the interface of software calibration


Impact direction should be set to 【 】 .
Five points should be measured vertically down on the
Leeb hardness test block.