Operation 
1.

Connect the steriliser to an alternating current supply of voltage specified on the rating plate. 

 
DANGER: THIS STEAM STERILISER MUST NOT BE CONNECTED TO A DIRECT POWER SUPPLY 
 
2.

Turn the simmerstat up as required and switch the MAINS on and check the indicating lamp

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

is illuminated. 

3.

An indicating lamp next to the simmerstat will illuminate when the heaters are operating. 

 
Maintenance 
The cabinet is finished with stainless steel and to maintain appearance should be wiped over with a

 

   

 

 

 

 

   

 

 

 

 

 

 

   

non abrasive cleanser. 
The interior is also stainless steel and may be cleaned with a solvent. 
The false bottom can be removed to allow cleaning around the element and controller sensor. 
A fixed gabled lid is fitted and the hinge on the back should be kept clean and free from foreign

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

objects to maintain easy opening. 
 
Troubleshooting 

SYMPTON 

REMEDY 

No Power 
 
(Indicator Light is off) 
 
 
 
 
 
 

1. Check steriliser is plugged in and power

 

   

   

 

 

switched on. 
 

2. Ensure mains power supply point is

 

 

 

 

 

 

functioning by using a test appliance on

 

 

   

 

 

 

power socket. 
 

3. Check internal RCD has not tripped. 

 

 

Failure to heat or maintain temperature 
 
(Indicator light is on) 

1. Ensure simmerstat is turned on 

 

2. Ensure the manual reset button is

 

 

 

 

 

 

depressed. 

 
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the

 

 

 

 

 

 

 

 

   

 

 

 

 

 

   

 

steam steriliser from the manufacturers label. 
 
 
Declaration of Conformity 
Each product is thoroughly inspected and tested to not only ensure that it meets the specifications 
provided, but to also meet Australian Electrical Standard AS3820 and EMC Standard 
AS/NZ1044:1995,and therefore being accredited with a C Tick label. 
 

3 Laboratory Equipment Pty Ltd Instruction Manual – Steam Steriliser 2013 ver 1.0 
 