Introduction 

Congratulations on the choice of an Australian made quality product. Labec products are manufactured, tested and

 

 

 

     

 

 

 

 

 

 

 

 

 

 

calibrated to meet published standard specifications under our strict quality assurance guidelines. 
 
This Instruction Manual is for the guidance of operators of Labec Water Baths and should be read before the water

 

 

   

 

 

   

   

 

 

 

 

   

 

 

 

 

bath is connected to the electricity supply. 
 
It is hoped that this manual will supply all the information that the customer should require for satisfactory

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

operation of the water bath. If, however, there are any questions that remain unanswered then the customer should

   

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

contact our service department. 
 
Unpacking 

Remove all packing and protective wrapping from both interior and exterior of the water bath. Check the water bath

   

 

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

for any possible transit damage. Ensure all ordered accessories are present. If any physical damage or shortages

 

 

 

 

 

   

 

 

 

   

 

 

   

 

are evident, do not discard the packaging material until the water bath is inspected by the distributor, agent or

 

 

 

 

 

 

 

 

 

 

 

   

 

 

 

 

   

manufacturer. 

NOTE: All claims for shortage or damage must be made within fourteen days (14) from delivery. 

Subject to our standard published conditions of sale, we have reasonable grounds to believe that we have ensured,

   

 

 

 

   

 

 

 

 

   

 

 

 

 

 

so far as is reasonably practical, that the products listed in our catalogue and brochures have been designed and

 

 

   

 

 

 

 

 

   

 

 

 

 

 

 

 

 

constructed so as to be safe and without risk to health when properly installed and used in their environment by

 

 

   

 

 

 

 

   

 

 

 

 

 

   

 

 

 

appropriate and trained personnel, and where applicable, in accordance with our published instructions. 
 
Installation 

Electrical 

This equipment must be tagged and tested according to AS/NZS3760:2010 prior to use and thereafter on

 

 

 

 

 

 

 

   

 

   

 

 

 

 

a regular basis dependent upon the environment.  

 
It is preferable to locate the bath close to a power point and recommended that double adaptors are not used.

   

   

 

 

 

     

 

 

 

 

 

 

 

 

 

 

Check the total wattage if connecting to multi-point outlets. Check the rating plate for power requirements.

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

Installation is to be carried out by a qualified electrician in accordance with the power requirements of the product

     

 

 

     

 

   

 

 

 

 

   

 

 

specifications. 

Ensure the unit is Tag Tested prior to first power up. 

 
Location 

Select a location free from draught and away from direct sunlight or other heat source. 
 
Circulating Pump 

For the circulating pump fitted there will be two small holes at the end of the bath. The upper hole is the inlet and

 

 

 

 

 

 

 

 

 

 

   

 

   

 

 

 

 

   

 

 

 

sucks the water in and then pushes it out the lower outlet hole across the heating elements under the tray. Ensure

 

 

   

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

before the bath is ever turned on that the upper inlet holes are well covered at all times. Test the water pump is

 

 

   

 

 

 

 

 

 

 

 

 

 

     

 

 

 

 

   

circulating with your finger and if not you will need to push some water into the inlet hole using a hose or similar to

 

 

 

 

   

 

 

 

   

 

 

 

 

 

 

 

   

   

   

push the airlock out of the pump system. 
 
Temperature Control 

LABEC water baths are fitted with either digital temperature controller as per Appendix A, connected to the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

   

 

element and cooling system. The water bath has been stringently tested before shipment to ensure all is in

 

 

 

 

 

 

 

 

 

 

 

 

   

 

     

working order. 
 
Description of Controls 

Set the desired temperature or heating level by adjusting the temperature controller as per Appendix A. A light on 

the controller “OP1” will illuminate when power is being supplied to the elements. Set the Alarm AL4 in Level 2 of 

the controller above the maximum temperature required about 5ºc above. Turn the cooling off above 30ºc. 
 
Cooling System 

The Cooling system uses a CFC free refrigeration compressor to cool the water in the bath. Turn if off above 30ºc 

to reduce wear on the cooling system. The cooling will run continuously and the heaters will maintain the 

temperature on the set value. Use “anti-freeze” when operating below 5ºc. Clean the cooling system air filter inlet 

monthly by wiping with a damp cloth or vacuum cleaner. 
 