LABEC PLRGD-1400PT User Manual

Page 3

Advertising
background image

4. Set the temperature controller at the required temperature using the up and down arrows on

 

 

 

   

 

 

 

 

   

 

 

 

 

the keypad. See Appendix A for more details. 

5. Final adjustment of the temperature controller may be required after the chamber has

 

 

 

 

 

 

 

 

 

 

 

 

 

reached operating temperature and this should be checked with a suitable thermometer

 

 

 

 

 

 

 

 

   

 

 

located in the chamber (A thermometer is not supplied with the chamber). 

6. The shelves can carry around 20 kilograms each. 

 
Temperature Controls 
See Appendix A 
 
Maintenance 
The cabinet is finished with stainless steel and to maintain appearance, should be wiped over with a

 

   

 

 

 

 

   

 

 

   

 

 

   

non abrasive household cleanser. The chamber is manufactured from stainless steel and may be

 

 

 

 

 

   

 

 

 

 

 

 

 

cleaned with a gentle cleaner. The door is also glass lined and should only be cleaner with general

 

   

 

 

 

   

 

 

 

 

 

   

 

 

 

purpose cleaners. 
The fan motor is pre lubricated with sealed bearings which should not require maintenance for some

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

time depending upon extent of usage. The shelves are manufactures from stainless steel and a non

 

 

 

   

 

 

 

 

 

 

 

 

   

 

abrasive cleaner should be used. The magnetic door gasket should be cleaned with detergent only,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ensuring it is dried completely after washing. 
 
Defrosting 
You can manually defrost by turning the unit off for a period of 24 hours with the door opened to

 

 

 

 

 

 

 

 

 

   

   

 

 

 

 

 

   

allow the frost to melt away. Otherwise the controllers are set to manual defrost. A defrost pan

 

 

   

 

 

 

 

 

 

   

 

 

 

 

 

under the unit will switch on daily to evaporate the condensation from the pan. 
 
Trouble Shooting 
 

SYMPTON 

REMEDY 

No Power 
 
(Indicator Light is off) 
 
 
 
 
 
 

1. Check fridge/freezer is plugged in and

 

 

 

   

 

power switched on. 
 

2. Ensure mains power supply point is

 

 

 

 

 

 

functioning by using a test appliance on

 

 

   

 

 

 

power socket. 
 

3. Check internal RCD has not tripped 

 

 

Failure to maintain temperature 
 
(Indicator light is on) 

1. Ensure the temperature controller set

 

 

 

 

 

point is above ambient. 
 

2. Check the safety controller fitted is above

 

 

 

 

   

 

the main controller setting. 

 
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the

 

 

 

 

 

 

 

 

   

 

 

 

 

 

   

 

unit from the manufacturers label. 
 
Declaration of Conformity 
Each product is thoroughly inspected and tested to not only ensure that it meets the specifications

 

   

 

 

 

   

 

 

 

   

 

 

 

provided, but to also meet Australian Electrical Standard AS3820 and EMC Standard

 

 

 

 

 

 

 

 

 

 

 

 

AS/NZ1044:1995,and therefore being accredited with a C Tick label. 
 

2        Laboratory Equipment Pty Ltd – Pass Through Fridges Instruction Manual 