Cleaning & disinfection / inspections / warranty, Cleaning & disinfection, Inspections & maintenance – Oakworks Marina User Manual

Page 13: Warranty

Advertising
background image

10

CleAnInG & DISInfeCTIOn / InSPeCTIOnS / WARRAnTy

CleAnInG & DISInfeCTIOn

Use a 10% sodium hypochlorite (bleach) solution or Recommended Disinfectants on all surfaces. Clean all sides

of the upholstered top & drape section. Follow the directions on the disinfectant and wipe off excess.
Recommended Disinfectants
Protex, MadaCide, Accell TB, Virox®

Note: Damage caused by unapproved substances will not be covered under the warranty.

DO NOT use citrus based cleaners or other strong cleaners, such as alcohol, acetone, higher concentrations of

bleach or other products that contain high concentrations of these substances.
DO NOT expose the fabric to temperatures below 50°F/10°C or above 104°F/40°C.
DO NOT expose the fabric to direct sunlight, adhesives, liquids, or abrasive materials.

InSPeCTIOnS & MAInTenAnCe

RECOMMENDED REGULAR INSPECTIONS (monthly or local standard)

• Check for damage to the power, hand control or foot control cables.
• Visually inspect components for obvious damage that could cause problems during operation.
• Base & Wood Trim: Use a mildew cleaner with soft brush as needed to prevent build-up of mold in

humid environments. Be sure excess cleaner does not drip onto, or into, any of the electrical controls or

mechanisms. Remove excess cleaner with a soft cloth.
• Inspect the fitted drape regularly for mold and mildew. Remove and air dry daily.

RECOMMENDED PERIODIC INSPECTIONS (yearly or local standard)

• Check for damage to the power, hands control or foot control cables and all visible wiring.
• Visually inspect components for obvious damage that could cause problems during operation.
• Check all mechanical functions using the hand control. Repeat using the foot control. Check for abnormal

noises.
• Replace any missing or illegible labels.
• Check that all fasteners are present and fastened securely.
• Check table grounding.
• Clean unusual buildup of dirt on the chair and/or parts of the chair not normally cleaned on a regular

basis.
• Check for tears or cracks in the upholstery.
• Inspect the Aqueduct and the drain periodically to remove any build-up of material that might interfere

with the flow of the treatment solution.

MAINTENANCE

• Follow the preventative steps under the Battery Use section of the Manual.
• Oakworks medical tables/chairs are designed and built to provide many years of dependable service.

Please follow local regulations and laws when disposing of the electrical components used in this chair.

For all Customer Service related problems refer to the Service Manual