Type 2 commands – balance input functions – Scientech 12000 Series User Manual

Page 17

Advertising
background image

17

Command Response

Purpose

[decimal number]_PCT[CR]

None

Puts the balance on the percent weighing mode with the
displayed reading converted to the percent entered. The
display must be non-zero when this command is issued. If
the display is zero, a 3 second "Err" message is displayed
and the balance returns to its normal weighing mode.
CLEAR exits the percent mode and un-defines the percent
register.

[Decimal number]_ENTER_[integer]
_PIECES[CR]

None

Accepts the decimal number as a weight in "last front panel
units" per integer pieces. If the balance is not in PCS mode,
it goes into PCS mode using the decimal number/the integer
as the weight. If used when the balance is already in PCS
mode, the display digits are blanked for about 1 second
while the PCS and "last front panel units" annunciators are
lit. The balance then resumes normal PCS operation using
the decimal number/the integer as the unit weight.

[register
number]_RCL_[(CONVERT) (CAL)
(PIECES)][CR]

REG XXX :[register
contents][MULT.]

Puts the recalled register contents into the specified mode
register and enters that mode.

TAVG[CR]

None

Clears the TBAR registers and enters the TBAR mode.

RESET_TAVG[CR]

None

Clears the TBAR registers. If the balance is in the TBAR
mode, the process keeps running.

XAVG[CR] None

Accumulates the current weight in the XBAR registers.

RESET_XAVG[CR] None

Clears the XBAR registers.

GLP[CR]

Sends GLP header

Puts balance in the GLP mode.

XAVG[CR]

Sends weight while in GLP

Accumulates the current weight in the XBAR registers.

RCL_XAVG[CR]

Sends statistics and signature
line

Terminates GLP mode.

ISO[CR]

Sends ISO header and waits
for target weight entry

Puts balance in ISO mode.

xxx.xx_Targ[CR]

Sends target weight (g)

Enter target weight value in grams.

SEND[CR]

Sends weight, difference
between weight and target,
and signature line

Terminates ISO mode.

ASON[CR] or ASOFF[CR]

See the RS-232 Balance
Message Formats section.

Turns auto send on or off. When auto send is on, a time
interval must be entered into register 89 (see Type 4
commands).

CSON[CR] or CSOFF[CR]

See the RS-232 Balance
Message Formats section.

Turns continuous send on or off.

ACAL[CR]

None

Starts the automatic internal calibration weight calibration
routine.





Type 2 Commands – Balance Input Functions:

Command Response

Purpose

[register number]_STORE[CR]

None

Stores the current weight, at the greatest possible resolution, in the
specified register.

[Decimal number]_ENTER_[register
number]_STORE[CR]

None

Stores the number entered, in the current units, as an additive
constant in the register specified.

[decimal number]_ENTER_CONVERT_[register
number]_STORE[CR]

None

Stores the number entered, as a "last front panel units" relative
multiplier, in the specified register.

[decimal number]_TARE[CR]

None

Adds the number entered, in current units, to the TARE register.
CLEAR [CR] zeroes the TARE register.

[decimal number]_ENTER_86_STORE[CR]

None

Stores the desired precision of the displayed weight for

T

mode.