OPERATION (CONTINUED)

S

ET THE

T

EMPERATURE

S

ET

P

OINT

Perform these steps to prepare the unit for calibration.

The unit comes from the factory set to 37°C

1. Turn the Over Temperature Limit control dial clockwise to the maximum position

indicated by the largest dot. This prevents the Over Temperature Limit system
from interfering with the Set the Temperature Set Point and Temperature
Calibration procedures.

2. Press either the Up or Down key one time on the Temperature Control panel to activate

the temperature set point mode. The temperature display will briefly flash the

letters “SP” to

indicate a Set Point is about to be displayed. The Set Temperature digital display will then
dim and start to blink, showing the adjustable temperature set point.

3. Use the Up and the Down key to adjust the set point

to your application’s or procedure’s

set point. If neither key is pressed within 5 seconds, the Set Temperature display will stop
blinking and return to displaying the current temperature of the incubator.

4. Wait 5 seconds after entering your set point. The display will stop flashing, and the set

point is now saved in the microprocessor controller. The incubator will now heat or cool
automatically adjust to match your set point.

5. Leave the Over Temperature Limit control at the maximum setting if you will be calibrating

the temperature display as part of the Preparing the Incubator procedure.