Shellab SMO10HP-2 User Manual

Page 12

Advertising
background image

12

at -2 to match the process value to the independent thermometer. To
navigate the control to the Offset Calibration mode, push the PGM button

once and then push the up arrow

once and the bottom display

should read CONF. Then push the PGM button again and the display

should say INP. Push the PGM button

again and the display should

read INP1. Push the PGM button

again and the display should read

SENS. This enters you in the Input/Configuration Menu. To navigate
through the menu, press the up and down arrows.

Press the up or

down arrow

repeatedly until you reach OFFS in the bottom display.

Then, push the PGM button

once, and the display will start flashing.

While flashing, use the up or down arrow

to set the desired Offset.

After the display stops flashing, push the Exit Button

repeatedly until

the NORMAL Display is reached.

6.7 Setting High Limit: Prior to setting the High Limit Control be certain the

Main temperature Control has reached the input set point and has
remained stable for several hours. This will insure that the High Limit is
set at the proper temperature. As stated in 4.3, the High Limit control
should be at maximum position, clockwise. Now turn the thermostat
counter-clockwise using a flat-head screwdriver until the High Limit
Activated indicator light comes on. Next turn the thermostat clockwise
just until the indicator light turns off. Then turn the thermostat clockwise
two (2) minor increments on its scale past the point where the indicator
light went out. This will set the High Limit at approximately 10

C above

the Main Temperature set point.