PULSER: By inhibiting heat build-up in the sample, the pulse function enables safe
treatment of temperature sensitive samples at high intensity. In addition, pulsing
enhances processing by allowing the material to settle back under the probe after
each burst. The ON and OFF pulse duration can be set independently from 1 second
to 59 seconds. During the OFF portion of the cycle, the red indicator on the PULSER
key will illuminate. If the OFF portion of the cycle exceeds three seconds, a cautionary
message – CAUTION – PROBE ON STANDBY – will warn the operator against
touching the ultrasonic probe. To set the pulser, press the PULSER key.

The screen will display:

Using the numeric keys set the ON portion of the cycle, then press the
ENTER/REVIEW key.

The screen will display: e.g.

Using the numeric keys set the OFF portion of the cycle.

The screen will display: e.g.

Press the ENTER/REVIEW key.

The screen will display:

Pulse on __.__sec
Pulse off __.__sec

Pulse on 01 sec
Pulse off __.__sec

TIME 5:30:25

TEMP __ __°C

PULSE 01 01

AMPL

40%

On Cycle

Off Cycle

Pulse on 01 sec
Pulse off 01 sec