SONICS Model

PSKIT-H540

This Sonics & Materials manufactured
ultrasonic kit is designed for

use as

an internal electrical component.
It should only be powered after being
properly mounted and earth grounded
inside of an industry standard
NEMA type electrical
enclosure.

IMPORTANT INFORMATION

1. NEMA enclosure needs to be properly fan cooled and vented.

2. Incoming line voltage (120 VAC 50/60 Hz.) must be noise filtered and

correctly earth grounded.

3. Because of potential noise interference, all RF signal cables and cable

connections should be routed as far away from other electrical
components as possible.

4. RF cable out from the NEMA enclosure and direct to a single ultrasonic

converter should not exceed 15’ (4.6 meters) in length.

5. When sequencing, RF cables out from the NEMA enclosure to individual

ultrasonic converters should not exceed 15’ (4.6 meters) in length.

6. To avoid RF cable capacitance variations, all RF cables should be equal

in length. Coil and zip-tie any excess RF cable lengths.

Description

Detail

Power Requirement

120 VAC 50/60 Hz 15 AMPS

Frequency

40 kHz

Power

500 Watts

Current Main Board Configuration

Sonics #855-00098 Rev 4

Current PLL Board Configuration

Sonics #855-00097 Rev 4