ThermoProbe TP7C User Manual

Thermo probe, Report of test, Calibration data as found

Advertising
background image

Model:

TP7C
7-15001

Date:
Calibrated By:

MC

Ambient Temp: ______+/- 3°F

Calibration Data As Found

In

Tolerance

º F

º C

º F

º C

º F

º C

º F

º C

º F

º C

If no "As Found" Data is shown above the Unit is new or no data was available

Calibration Data As Left

This device has been adjusted to read as closely as possible to actual temperature.
Tested temperatures and corrections are as follows:

In

Tolerance

º F

º C

º F

º C

º F

º C

º F

º C

º F

º C

32

0.0

32.0

0.00

31.9

-0.06

0.1

0.06

0.2

0.11

Yes

120

48.9

120.0

48.89

120.0

48.89

0.0

0.00

0.2

0.11

Yes

199

92.8

199.0

92.78

199.0

92.78

0.0

0.00

0.2

0.11

Yes

Date:

300

149.0

300.2

149.00

300.1

148.94

0.1

0.06

0.5

0.28

Yes

The results stated on this report relate only to the items specifically identified.

This test report or calibration certificate shall not be reproduced except in full,

Note: Celsius values converted from Fahrenheit.

without written approval of the laboratory.

Tolerance

Correction

Reading of TP

Actual Test

Temp.

Correction

Unit SN:

Nominal Value

Nominal Value

6/3/2013

6/3/2013

Reading of TP

Actual Test

Temp.

Tolerance

Report of Test

This report is to certify that the
instrument listed below has been
calibrated by ThermoProbe, Inc. to
NIST traceable criteria.

Traceability: This calibration is traceable to NIST through an unbroken chain of
comparisons. Thermoprobe calibrated the references shown above using a
Secondary Standard (SSPRT) which in turn was calibrated by Isotech using
their primary reference thermometer.


Calibrator's Signature: ___________________________________________


Test Results Approved by: _______________________________________


Test Method: The calibration procedures used were ThermoProbe, Inc. Calibration Procedures
based on ASTM E-644-06. This probe was immersed in a constant temperature bath with a
reference thermometer which determined the actual test temperature. The readings were
compared and correction factors for the probe were calculated. The As Left readings reflect the
thermometer's readings after calibration.

Report of Test

ThermoProbe

Reference

Calibration Date

Next Calibration Due

32.0º F

0.0º C

TL1R - 5169

12/19/2012

12/19/2013

120.0º F

48.9º C

TL1R - 5769

12/19/2012

12/19/2013

199.0º F

92.8º C

TL1R - 4849

7/19/2012

7/19/2013

300.2º F

149.0º C

TL1R - 4850

7/19/2012

7/19/2013

Nominal Temp