Use and operation

Installation:

Max. ambient temperature: 40 °C

Remove the protective foils from the fittings at the flask head and the protective
covers from the inlet of the diaphragm pump or the solenoid operated valve.

Assemble the connecting tube with filter between the bottle head and the inlet of the
diaphragm pump (BVC 21) or the inlet of the solenoid operated valve (BVC 01).

Connect the aspiration tube of the VacuuHandControl VHC to the coupling at the
bottle head.

Connect the female connector of the mains cable to the pressure control device.

Make sure ventilation is adequate if pump is installed in a housing or if ambient
temperature is elevated. Keep a distance of min. 20 cm between fans and ambient
parts.

If pump is installed in altitudes of more than 1000 m above mean sea level check
compatibility with applicable safety requirements, e. g. IEC 60034 (motor may
overheat due to insufficient cooling).

Before switching on the pump check that mains voltage and current conform
with the equipment!

Pump with dual-voltage motor:

Check that the voltage changeover switch underside the terminal box is posi-
tioned correctly.

Attention: If the pump is switched on with the voltage not correctly selected,
the motor may be damaged!
Pay attention to the warning label on the on/off
switch and remove after having set the voltage changeover switch to the correct
position!

Selecting the supply voltage:

Attention:
Before starting the pump, check that
the voltage selection switch under-
side the terminal box is correctly
positioned.
Change the selection at the voltage
changeover switch only if the pump
is separated from mains.

Voltage selection switch:
Use a screw driver to adjust the supply
voltage on the voltage changeover
switch
at the terminal box of the pump:
”115” corresponds to 100-120V and
”230” corresponds to 200-230V.

During operation

:

The pump achieves its pumping speed, ultimate total vacuum and vapour pumping
rate only at operating temperature (after approx. 15 minutes).

Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves
will be damaged, if liquids are transfered through the pump in significant amounts.