Prior to autoclaving set the VHC

pro

to

“autoclaving”.

Prior to autoclaving loose the cap.

Cleaning and decontaminating

Autoclaving

The VacuuHandControl VHC

pro

with adapters

and aspiration hose is designated for steam

sterilization at 121°C and 2 bar absolute (1 bar

overpressure). Time of exposure according to

DIN 58946 t

e

= 20 minutes.

Ensure not to exceed a temperature of 121°C.

Comply with instructions for use of the manu-

facturer of the autoclave.

It is the user´s responsibility to ensure effective

autoclaving.

NOTICE

NOTICE

After autoclaving tighten the nose cones if nec-

essary with the mounting tool (enclosed in de-

livery), see section „Replacing the nose cones“.