Montage of the connecting hose:

If the pump does not achieve the ultimate pressure:

☞ In case the diaphragms and valves have been replaced, a run-in period of several

hours is required before the pump achieves its ultimate vacuum.

If all pump heads achieve a vacuum below 120 mbar but pump does not achieve the
ultimate total pressure:
Check hose connectors between pump heads and manifolds for leaks. If necessary
recheck pump chamber.

☞ Reassemble the pumping unit.

➨ Use open ended wrench (w/f 15) to reconnect hose to elbow

fitting.

➨ Tighten union nuts first by hand and then tighten one full

turn using open ended wrench (w/f 17).

Changing the valve at the outlet manifold

➨ Note position of valve and remove.

☞ Check valve for damage and replace if necessary. Make

sure that the valve is correctly seated.

➨ Reassemble cover plate and manifold cover.

➨ Use open ended wrench (w/f 15) to reconnect hose to elbow

fitting

➨ Tighten union nut first by hand and then tighten one full turn

using open-ended wrench (w/f 17).

➨ Use open-ended wrench (w/f 17) to loosen the union nut of

the hose, which runs directly to the cover plate of the outlet
manifold, at the pump head.

➨ Use open-ended wrench (w/f 15) to turn elbow fitting 1/4 of a

turn, remove hose. Do not remove the elbow fitting from the
pump head.

➨ Unscrew the countersunk head screws at the manifold cover

and remove cover plate and manifold cover.