VACUUBRAND MZ 2C +AK+EK User Manual

Page 16

Advertising
background image

page 16 of 28

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 995653 / 06/10/2009

Removing the catchpots:
Catchpot at outlet:
➨ Remove joint clip, remove catchpot and drain condensate.

Catchpot at inlet:
➨ Admit air or inert gas (via inlet of pumping unit) to atmospheric pressure. Remove

joint clip, remove catchpot and drain condensate.

➨ Reassemble drained catchpots.
Important: Comply with regulations when disposing solvents/condensates. Reuse if
possible, purify if contaminated.

Shutdown:

Short-term:
Has the pump been exposed to condensate?
☞ Allow the pump to continue to run at atmospheric pressure with gas ballast valve

open for a few minutes.

Has the pump been exposed to media which may damage the pump materials or forms
deposits?
☞ Check and clean pump heads if necessary.

Long-term:
➨ Take measures as described in section short-term shutdown.

➨ Separate pump from the apparatus.

➨ Close manual gas ballast valve.

➨ Drain catchpots.

➨ Close inlet and outlet port (e. g. with transport caps).

➨ Store the pump in dry conditions.