Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999122 / 02/09/2009

page 14 of 21

To check the valves, remove head cover carefully from hous-
ing cover.
➨ Use hex key to remove four socket-head screws from pump

head and remove upper housing (head cover with housing
cover and valves).

☞ Never remove parts by using a spiky or sharp-edged tool

(e. g. screw driver), we recommend to use a rubber mallet
or compressed air (to be blown carefully into port).

➨ Note position of valves and remove them.

☞ Replace valves if necessary.

Cleaning and inspecting the pump heads:

➨ Use open-end wrench (w/f 17) to remove screw-in fittings

from all pump heads and remove manifold. Steady fitting
with a second open-end wrench (w/f 20).

➨ Detach manometer from adjustment device using open-end

wrenches w/f 14 and 17.

➨ Unscrew hollow bolt at adjustment device with open-end

wrench w/f 17 whilst steadying fitting with a second open-
end wrench (w/f 20).

☞ Check diaphragm for damage and replace if necessary.

➨ To do so use Phillips screw driver to remove four counter-

sunk head screws and lift off housing plate.

☞ Use petroleum ether or industrial solvent to remove depos-

its. Do not inhale.

Advertising