Ad604, Ib ^ il j – Analog Devices AD604 User Manual

Page 22

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

AD604

MEDICAL ULTRASOUND TGC DRIVING THE
AD9050, A 10-BIT, 40 MSPS ADC

The AD604 is an ideal candidate for the time gain control (TGC)

amplifier that is required in medical ultrasound systems to limit
the dynamic range of the signal that is presented to the ADC.

Figure 52 shows a schematic of an AD604 driving an AD9050
in a typical medical ultrasound application.

The gain is controlled by means of a digital byte that is input to

an AD7226 DAC that outputs the analog gain control signal.
The output common-mode voltage of the AD604 is set to VPOS/2

by means of an internal voltage divider. The VOCM pin is
bypassed with a 0.1 pF capacitor to ground.

The DSX output is optionally filtered and then buffered by

an AD9631 op amp, a low distortion, low noise amplifier. The

op amp output is ac-coupled into the self-biasing input of an
AD9050 ADC that is capable of outputting 10 bits at a 40 MSPS
sampling rate.

0.1MF

0.1MF

ANALOG

INPUT

500

-w.-----

O.iMFij::

-------- Mi

0.1

m

F

------ ----------

-DSXI

VGN1

+DSX1

vref

PAO1

OUT1

FBK1

GND1

PAII

VPOS

COMI

VNEG

COM2

VNEG

PAI2

VPOS

FBK2

GND2

PAO2

OUT2

+DSX2

VOCM

-DSX2

VGN2

AD6O4

_ 1Oo

0 ]

v

out

B

VoutA

IB

^ il J

V

OuT

C

VoutD

V

DD

.. AD7226

Vref

AO

AGND

A1

DGND

WR

DB7

DB0

(MSB)

(LSB)

DB6

DB1

DB5

DB2

DB4

DB3

AO

17|—I

A1

l

e

T-l

WR 1

5

1-4

I------

DIGITAL GAIN CONTROL

Figure 52. TGC Circuit for Medical Ultrasound Application

Rev. E | Page 22 of 32

J2

5O0

19

+15V

11

Advertising