40-4328

Weighing Indicator

Physical Dimensions

170/6.69"

AO-4328 WEIGHING INDICATOR

•<

88B888

lb

kg

ZERO MO OAOSS NTT PT

o o

o

Q© GD ©

fouSìì fìCToì Ì

t

«»

Front View

Side View

+1

Unit : mm / inches

162±0.5/6.38±0.02'

Cut Out

Specifications

Input Sensitivity -

Power

Input Signal Range

Loadcell Excitation

0.2p/d (Min.)

-ImV - +15mV

■5V DC ±5%

Loadcell Drive Capacity — Up to 6 load cells at 350fl/load cell

(Max.90mA)

Zero Temperature Coeffdent--- ±(0.2pV+0.0008% of dead load)/*C (typ.)

Span Temperature Coeffdent---±8ppm rc of reading (typ.)

■0.010/ooffuliscaie

±0.2pVp-p

• 10MQ ± or more

• int^ating dual-siope type

- 40,000 counts (Max.)

• Approximately. 10 times/second

- edigits, 14.2mm / 0.56" (H) LED

■ 20,000 counts (Max.)

Times XI, X2, X5, XI0, X20, X50

500000

-

minus display

- ZERO, MD, CROSS, NET,

PRESET TARE, kg/t or kg/lb

- STANDBY/OPERATE, NET/CROSS, PRESET

TARE, ZERO, M+, TARE, MODE, PRINT

Non-Linearity---------------

Input Noise------------------

Input Impedance----------

A/D Conversion Method

/V/D Resolution -----------

A/D Conversion Rate -

Weight Display--------------

Display Resolution —

Minimum Division ---------

Maximum Display----------

Under ZERO Indication
Annuciation------------------

Keys

Net Weight ----------------

Operating Temperature

Storage Temperature

Physical Dimensions

Panel Cut out Dimensions

Standard Accessories

Options

OP-01 —

OP-02 —

OP-03 —

OP-04 —

OP-10 -

-DC 9V

(AC Adaptor or direct input to the lermial)

- Approximately 0.8 kg /1.8 lb

--lO'C-AO'CiU'F- 104'F)

--10°C--70‘’C(14°F- 158 °F)

- 170(W) X 126 +19" (D) X 130(H) mm

•|ncl. terminal cover

-124 ±0.5 X 162 ±0.5 mm

- Manual, AC Adaptor

BCD Output (Open Collector)

Comparator Output (Open Collector)

RS-422/485 with Relay Output

RS-232C/Current Loop/Relay Output

Display Stand

Only one output can be selected from OP-01, OP-02, OP-03 and OP-04.

One to be choosen from RS-422 or 485.

Accessories

AC Adaptor —

TB164 (220V)

Specifications subject to change for improvement without notice.

A&O WEIGHING

...Clearly a Better Value

1555 McCandless Drive, Milpitas. CA. 95035

Toll-free (800) 726-3364

Tel:(408)263-5333 Fax:(408)263-0119 http>/www.and1.com

♦ADE-AD4328-961210-01KO1