3.5.3 Calibration of a Toxic Gas Channel

Equipment Required

‰

The calibration cover

‰

A calibration adapter for the type of gas to be calibrated:

ƒ

For reactive gases, 8” Teflon lined tubing, cylinder regulator and reactive gases (ex. H2S, NH3, SO2, HC1 etc…)

ƒ

For non-reactive gases, Tubing and cylinder regulator

‰

A cylinder of calibration gas with a known content and corresponding to the type of sensor to be calibrated

Preparing the OMNI-4000

• Attach the OMNI-4000 calibration cover. See Figure 3-8

• Connect the calibration gas cylinder to the calibration cover with a vinyl hose

• If the channel has not been selected refer to the "Selecting the Channel for Calibration" paragraph

Use:
Cylinder Regulator 02506-004 with 17 liter cylinder

Cylinder Regulator 02506-002 with 34 liter cylinder
(ex. H2S, NH3, HCL etc.)
Cylinder Regulator 02506-005 with 103 liter cylinder
(ex. CO2 )

Calibration Arrangement for the Majority of Gasses

Humidifier
Used with BRH Sensor

Fill humidifier bowl 1/3 full with clean, tap water. The tube
should

NOT

go under the water line. No bubbles should be

visible when gas is flowing
Cylinder regulator part number 03700-005.

Top View of Humidifier

From Regulator О

Н

To Sensor

Note: In this application, the direction of gas flow is the

opposite of the indicator arrow on the humidifier
assembly.

Calibration Arrangement for BRH Gases

Figure 3-8: Calibration Gas Equipment Arrangement

Calibration Cover

Tubing – Teflon lined

Gas Cylinder

Regulator

Preset @ 0.5 lpm

Humidifier

Calibration Cover

Regulator

Gas Cylinder