background image

ENMET Corporation

TX2000 / OX2000

14

4.4 Sensor Replacement

Necessary when:

• Instrument fails to calibrate, sensor is worn out (DEF4).

• Wrong type of sensor installed (DEF128).

N

OTE

: If other type of sensor is connected by mistake during sensor replacement, “DEF128” fault will be triggered.

Table 3: Data for Oxygen and Toxic Sensors

Gas /

Unit of measure

Standard range

Accuracy

(in % of full scale)

Resolution

Zero drift

Response time

(in seconds)

Expected Life of

Sensor

(in months)

Warranty

(in months)

CL2 ppm (2)

10

< 5 %

0.1 ppm

< 0.5 ppm

per 6 months

< 180

> 12

12

CO ppm

500

< 5 %

1 ppm

< 20 ppm

per 6 months

< 45

24

12

H2 ppm

2000

< 5 %

1 ppm

< 50 ppm

per 6 months

< 150

24

12

H2S ppm

100

< 5 %

1 ppm

< 1 ppm

per 6 months

< 30

24

12

HCN ppm

30

< 5 %

0.1 ppm

< 0.5 ppm

per 6 months

< 90

18

12

NH3 ppm

100

< 5 %

1 ppm

< 3 ppm

per 6 months

< 220

> 12

9

NO ppm

100

< 5 %

1 ppm

< 2 ppm

per 6 months

< 30

> 12

12

NO2 ppm

30

< 5 %

1 ppm

< 1 ppm

per 6 months

< 100

> 12

12

O2 %

by Volume

30

< 0.5 %

volume O2

0.1% vol.

< 2%

per month

< 20

> 12

12

O3 ppm (1)

1

± 0.05 ppm

0.01 ppm

< 0.05 ppm

per 6 months

< 120

12

12

N

OTE

: All specifications stated in this manual may change without notice.

(1) Require use of Sample Draw System. See Figure 5.
(2) Recommended use of Sample Draw System, for enhanced sensitivity and speed of response. See Figure 5.

Example of Sample Draw System
with Sampling Adapter 02552-023

Figure 5: Calibration Adapter with Sample Draw System

MOTORIZED SAMPLE

PUMP

OUT

IN

CHARGE

JACK

ALARM

POWER

Motorized Sample Pump

03700-029

Sample OUT
To Sensor

Ambient Air
Sample IN

Adapter with Fitting for Calibration