ENMET ISA-60M with MRI-5175 User Manual

Page 22

Advertising
background image

ISA-60M

ENMET Corporation

19

5.2.11 Exit Maintenance Menu

From the "

exit MAINTENANCE Menu

" position

Press the

SELECT

switch to resume the operational display.

Press the

OPTION

switch to reenter the maintenance menu at the "

Calibration

" position.

Example: Exit Maintenance Menu Flow Chart

5.3 MRI-5175 Sensor Transmitter

The

MRI-5175

is supplied with 3 alarm points:

2 are factory set to match the

ISA-60M

The 3

rd

is set at 23.5 as an oxygen abundance indicator.

The alarm points are not field adjustable.

When equipped with an audio alarm for safety reasons this alarm can not be silenced when triggered by an unsafe condition.

5.4 Channel Activation/Deactivation

Each channel of the

ISA-60M

can be deactivated and activated as needed.

C

AUTION

:

Deactivated channels will not display or respond to connected sensor transmitters.

A fault indication occurs when a transmitter is not connected to an active channel of the

ISA-60M

.

To Activate or deactivate a channel of the

ISA-60M:

Enter the maintenance menu as described in Section 5.2 and press the

OPTION

switch until Exit maint menu appears.

Press and Hold the

OPTION

switch until “

ADV maint menu”

is displayed.

Press the

SELECT

switch to locate the channel to be changed.

Press the

OPTION

switch to change the status of that channel.

Press the

SELECT

switch to accept the change in status. “

ADV maint menu Exit”

is displayed.

Press the

SELECT

switch to return to the operational menu.

S

O

MAINTENANCE MENU

Exit maint menu

MAINTENANCE MENU

Calibration

CH-1

CH-3

CH-2