Guide for MRI-5175 Sensor Replacement

An Oxygen sensor must be replaced when it can no longer be calibrated. Typical sensor life is 3 to 5 years in a 20.9%
oxygen environment.

To replace a sensor, perform the following steps:

Turn off the electrical power.

Open the display panel and remove the four cover (lid) retention screws and remove the lid. See Figure 1.

Remove the old O

2

sensor, by unscrewing it, and replace it with a new sensor. See Figure 1.

Replace cover(lid) with retention screws.

Turn on the electrical power.

Wait 3 – 4 hours, then recalibrate the sensor per the instructions.