4.3.2

4-20mA Current Sink

The Input Module can also sink current proportional to the detected gas level.

i.e.

zero gas

= 4mA

full

scale =

20mA

The supply is derived from the external equipment.

Notes

Switch ON J10 and J13

J10

J11

J12

J13

J14

NOTE: Incorrect jumper switch configuration can cause damage to the system.