ENMET EX-5130 User Manual

Page 11

Advertising
background image

EX-5130-IS

ENMET Corporation

9


5.2.1 Zero Adjust

A ZERO function should be performed only when the

EX-5130-IS

sensor/transmitter is exposed to fresh air. If the air

at the sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference. Attach the cylinder to the
calibration adapter.

Enter the maintenance menu by placing the magnet over

M

ENU

switch for 2 to 4 seconds. See Figure 5,

EX-5130-IS

Maintenance Menu flow chart.

The second menu available is the Zero.

Tap the

S

ELECT

switch to perform a Zero.

If the Zero is successful, Cal OK appears on the display and in 1 – 2 seconds, display will change to Span.
If you wish to Span the sensor Tap the

S

ELECT

switch you are now ready to apply gas. Proceed to gas span step 2

If you wish to Exit the maintenance menu, Tap

M

ENU

switch until Exit is displayed, then tap

S

ELECT

switch to

return to the instrument Normal Gas Display

If the Zero is Not successful, sensor is outside of safe parameters to be zeroed, the display will read Bad Zero.
Repeat Section 5.2.1 Zero Adjust making sure to use a cylinder of 20.9% Oxygen.

5.2.2 Gas Span

It is recommended that the Zero Function be performed first.

Enter the maintenance menu. See Figure 5,

EX-5130-IS

Maintenance Menu flow chart.

1.

Tap the

M

ENU

switch once to show Span on the display.

2.

Tap the

S

ELECT

switch to perform a Span procedure. The display will alternate between the calibration gas

concentration and a signal level.

3.

Attach the associated calibration gas cylinder to the regulator and calibration cover. See to Figure 3.

4.

Open the valve to apply the calibration gas to the sensor.

5.

Watch for the signal level to stabilize. Refer to Table 4 for typical response times.

6.

Once the signal level has stabilized, the

EX-5130-IS

will automatically lock in the calibration data and:

If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1. Remove
calibration gas. To exit maintenance menu tap the

M

ENU

switch until Exit appears, then tap the

S

ELECT

switch.

If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span.
Remove calibration gas. Tap the

M

ENU

switch until Exit appears, then tap the

S

ELECT

switch. Check span gas

and repeat calibration in 30 – 60 minutes.

If the sensor did not respond to gas, Same mV is displayed momentarily, then returns to Span.

Remove calibration gas, tap the

M

ENU

switch until Exit appears, then tap the

S

ELECT

switch and try

calibration again in 30-60 minutes.
If the sensor will not calibrate See Section 5.4.

N

OTE

:

Some software revisions require the

S

ELECT

switch be tapped to accept the signal.

7.

Calibration is complete.

5.2.3 Exit Maintenance Menu

Exit maintenance, by tapping on the

M

ENU

switch until Exit appears on the display. Tap the

S

ELECT

switch to return

to the instrument Normal Gas Display.

N

OTE

:

You can change the Calibration Gas Level. H

OLD

the magnet over the S

ELECT

switch for 2 – 4 seconds

The M

ENU

switch changes digit indicated by underscore cursor

The S

ELECT

switch locks underscored digit and moves to next digit

Advertising