Measuring Power or Loss (FPM-600 and FOT-600)

Power Meter/Light Source/Optical Loss Test Set

25

Referencing Your Power Meter to a Source

To reference the power meter to a source manually:
1.
Using the proper adapter, connect a light source (such as FLS-600 or

FOT-600) to the detector port of your power meter.

2. Activate the source at the desired wavelength.

3. Match the source and power meter

wavelengths:

If the source emits an auto-wavelength
signal or is in auto-switching mode (see
Automatically Detecting Wavelength on
page 29
and Using Auto-Switching Mode
on page 35), the power meter automatically matches the source
wavelength.

OR

Press

to switch between pre-selected wavelengths (see

Defining a List of Favorite Wavelengths on page 22).

4. Hold down

for a few seconds. The

power meter stores the currently detected
power as the new reference power.

Reference power is displayed in the top
right corner (in dBm) and current loss
reading is automatically switched to dB.

Light source

Power meter

Adapter

Reference

test jumper

Reference

test jumper

Bulkhead

connector