Troubleshooting

268

FTB-500

Solving Common Problems

Battery indicator shows only

a fraction of the power is

left, but the batteries are

fully charged.
(You are able to operate

your unit, but the software

automatically blocks when it

reaches its set power

threshold).

Battery needs to be

recalibrated or changed.

Use the battery

calibration process

explained in Equation

on page 254.

Change the batteries.

When I start the probe

application, an error

message warns me that the

probe application is not

installed on my unit.

The fiber inspection probe

application is no longer

installed on FTB-500 units by

default.

Install the fiber inspection

probe application

(ConnectorMax) using

Update Manager. For more

information on how to

install products, refer to the

Update Manager online

help. For more information

on how to use the

ConnectorMax application,

refer to the corresponding

online help.

When I start the probe

application

(ConnectorMax), the probe

is no longer detected.
OR
No image is displayed in the

probe application.

There may be a problem

with the probe’s driver.

Install the latest System

Upgrade kit on your unit,

then try again.

Problem

Possible Cause

Solution