Use a larger number of sops and wavelengths – EXFO FTB-5600 Distributed PMD Analyzer User Manual

Page 128

Advertising
background image

Optimizing Measurements

122

FTB-5600

Use a Larger Number of SOPs and Wavelengths

Use a Larger Number of SOPs and Wavelengths

This will ensure that the instrument can access close to a maximum
available wavelength number from the instrument. More SOPs can also
improve the measurement accuracy for RMS DGD in the available
wavelength range.

A maximum wavelength range and as many SOPs as possible for the
acquisition will reduce the uncertainty of a measurement (that is, to
achieve a minimum measurement uncertainty).

Note: At least 100 or 200 SOPs should be used for obtaining a meaningful

cumulative PMD curve as a function of distance, and more SOPs. For
example, 500 to 1000, is recommended in order to make a better and
clearer picture of cumulative PMD.