SCPI Command Reference

50

FTB-9100

Product-Specific Commands—Description

:ROUTe[1..n]:SCAN:ADJust:AUTO

Description

This command sets the <Optimize for

repeatability> option. When this option is

selected, the switch mechanism returns to a

reference position before aligning to a new

position. This ensures optimum repeatability.

When this option is not selected, the switch

mechanism goes directly to the new channel

position. This provides faster switching times, but

less repeatability.

On *RST, the value of this setting is OFF.

Syntax

:ROUTe[1..n]:SCAN:ADJust:AUTO<wsp><Auto

Adjust>

Parameter(s)

AutoAdjust:
The program data syntax for <AutoAdjust> is

defined as a <Boolean Program Data> element.

The <AutoAdjust> special forms ON and OFF

are accepted on input for increased readability.

ON corresponds to 1 and OFF corresponds to 0.

The <AutoAdjust> parameter enables or

disables the optimization for repeatability.

1 or ON, enables the optimization for

repeatability.

0 or OFF, disables the optimization for

repeatability.

Advertising