EXFO PM-1100 Optical Power Meter User Manual
Ddendum
P/N: 1051923
January 2008
1/4
A
DDENDUM
This information applies only to the
PM-1100 Optical Power Meter user guide.
GPIB Commands and Queries
³
The FORM:READ:DATA? query does not exist.
³
The MMEM:ACQ:DATA:RECA:VALUE? query should be defined as
follows:
MMEMory:ACQuisition:DATA:RECAll:VALUE?
Description This query returns the measurement saved in the <data>
memory location.
Syntax
MMEM:ACQ:DATA:RECA:VALUE?<space><data>
Parameters The <data> parameter represents the memory location where
measurement data is stored (out of the 1024 available).
Response
Measurement data saved in the specified memory location in
the “±999.999E±99” format.
Notes
³
Use the MMEM:ACQ:DATA:RECA:UNIT? query to get the
corresponding measurement units.
³
Use the MMEM:ACQ:DATA:RECA:WAVE? query to get the
wavelength at which the measurement was taken.
Example
MMEM:ACQ:DATA:RECA:VALUE? 1021