Remote Control

Tunable Laser Source

101

Specific Commands

SOURce:SWEep:REVerse

Description This command is used to enable and disable the sweep reverse

function. When this function is enabled, the sweep program is

performed in both directions: ascending and descending.

Syntax

SOUR:SWE:REV<wsp><sweep>

Parameters The <sweep> parameter is a boolean parameter indicating

whether or not the sweep will be performede in both

directions:

1: enable the sweep reverse function.

0: disable the sweep reverse function.

Example

SOUR:SWE:REV 1

Note

This command cannot be performed during a sweep.

See also

SOUR:SWE:MODE, SOUR:SWE:MODE?, SOUR:SWE:CENT,

SOUR:SWE:COUN, SOUR;SWE:COUN?, SOUR:SWE:REP,

SOUR:SWE:REP? and SOUR:SWE:REV?

SOURce:SWEep:REVerse?

Description This query returns a value indicating whether the sweep

reverse function is enabled.

Syntax

SOUR:SWE:REV?

Response

A boolean parameter representing the state of the sweep

reverse function:

1: the sweep reverse function is enabled.

0 : the sweep reverse function is disabled.

See also

SOUR:SWE:MODE, SOUR:SWE:MODE?, SOUR:SWE:CENT,

SOUR:SWE:COUN, SOUR;SWE:COUN?, SOUR:SWE:REP,

SOUR:SWE:REP? and SOUR:SWE:REV