Dead zone calibration – EXFO IQS-12002B Calibration System User Manual

Page 81

Advertising
background image

Calibration Methods

Calibration System

77

FTB-7000D and FTB-7000B-B OTDR Modules

Dead Zone Calibration

The dead zone depends on the intensity of the reflectance. Compliance to
the specification must be done for a reflectance of –35 dB in multimode
and –45 dB in singlemode.

Note: There is no uncertainty value associated with this calibration.

Attenuation Dead Zone

The attenuation dead zone is the distance between markers A and B
on the OTDR trace presented below.

Marker A is positioned at the rising edge of the reflective event.
Marker B is positioned within the falling edge, 0.5 dB above the
backscatter level.

Attenuation
dead zone

Advertising