Hanna Instruments HI 991003 User Manual

Instruction manual, Portable ph/mv meter with sensor check

Advertising
background image

Dear Customer,
Thank you for choosing a Hanna Instruments
product.
Please read this instruction manual care-
fully before using the instrument. If you need
additional technical information, do not hesi-
tate to e-mail us at [email protected]
This instrument is in compliance with the

directives.

All Hanna Instruments meters are warranted
for two years
against defects in workmanship
and materials when used for their intended pur-
pose and maintained according to instructions.
The probes are warranted for a period of
six months.

This warranty is limited to repair or replacement
free of charge. Damage due to accidents, mis-
use, tampering or lack of prescribed maintenance
are not covered.

If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of pur-
chase, serial number and the nature of the prob-
lem. If the repair is not covered by the warranty,
you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instru-
ments, first obtain a Returned Goods Authoriza-
tion Number from the Customer Service depart-
ment and then send it with shipment costs pre-
paid. When shipping any instrument, make sure it
is properly packaged for complete protection.

All rights are reserved. Reproduction in whole or
in part is prohibited without the written consent of
the copyright owner.

WARRANTY

PRELIMINARY EXAMINATION

Remove the instrument from the packing
material and examine it carefully to make
sure that no damage has occurred during
shipment. If noticeable damage is evident,
notify your dealer.

Note: Save all packing material until you

are sure that the instrument func-
tions correctly. All defective items
must be returned in the original
packing together with the supplied
accessories.

HI 991003 is a portable pH/pH-mV/ORP/
temperature meter, with unique Sensor
Check feature, that allows the user to de-
termine the electrode status at any time.
Main features also include: extended mea-
surement ranges; waterproof and compact
casing; large dual-line LCD; low battery de-
tection; automatic pH calibration at one or
two points with two memorized buffer sets;
selectable temperature unit (°C or °F).
The pH/ORP electrode has a built-in tem-
perature sensor for simultaneous pH (or
mV) and temperature readings, and also
contains a mini amplifier to render mea-
surements impervious to noise and elec-
trical interferences.
Each meter is supplied complete with

HI 1297D pH/ORP/temperature probe with

DIN connector and 1 m (3.3’) cable

calibration & cleaning solution sachets

3x1.5V AA alkaline batteries

instruction manual

rugged carrying case.

GENERAL DESCRIPTION

SPECIFICATIONS

Range (*)

-2.00 to 16.00 pH

±825 mV (pH-mV)

±1999 mV (ORP)

-5.0 to 105.0°C / 23.0 to 221.0°F

Resolution

0.01 pH

1 mV

0.1°C / 0.1°F

Accuracy

±0.02 pH

(@20°C/68°F)

±2 mV

±0.5°C up to 60°C; ±1°C outside

±1°F up to 140°F; ±2°F outside

Typical EMC

±0.02 pH

Deviation

±2 mV

±0.2°C or ±0.4°F

Temperature

Automatic for pH readings

Compensation

pH Calibration

Automatic, 1 or 2 point

with 2 sets of memorized buffers

(pH 4.01/7.01/10.01 or 4.01/6.86/9.18)

mV Calibration

Factory calibrated

Probe

(included)

HI 1297D

amplified pH/ORP/temperature

probe with DIN connector and 1 m (3.3’) cable

Battery Type

3 x 1.5V AA / IEC LR6

Battery Life

Approx. 1500 hours

Auto-off

After 8 minutes of non-use

Environment

0 to 50°C (32 to 122°F);

RH max 100%

Dimensions

150 x 80 x 36 mm (5.9 x 3.1 x 1.4”)

Weight

210 g (7.4 oz.)

(*)

the temperature range is limited to 80°C (176°F) if

using the HI1296D or HI1297D probes.

To clean the meter, use water only.

HI 1297D

Combination amplified pH/
ORP/temperature probe with
DIN and 1 m (3.3’) cable

HI 1296D

Combination amplified pH/
temperature probe with DIN
connector and 1 m (3.3’) cable

HI 7004L

pH 4.01 buffer solution, 500 mL

HI 7006L

pH 6.86 buffer solution, 500 mL

HI 7007L

pH 7.01 buffer solution, 500 mL

HI 7009L

pH 9.18 buffer solution, 500 mL

HI 7010L

pH 10.01 solution, 500 mL

HI 7020L

200/275 mV test solution, 500 mL

HI 70300L

Storage solution, 500 mL

HI 7061L

Cleaning solution, 500 mL

HI 710007

Blue shockproof rubber boot

HI 710008

Orange shockproof rubber boot

HI 721312

Rugged carrying case

HI 76405

Electrode holder

ACCESSORIES

LCD DESCRIPTION

IST991003R1 07/05

1. Automatic Temperature Compensation
indicator
2. Stability indicator
3. Battery percentage
4. pH calibration indicator
5. Low battery indicator
6. Selectable temperature unit
7. Secondary display
8. Primary display
9. Measuring unit for primary display

w w w . h a n n a i n s t . c o m

Instruction Manual

HI 991003

Portable pH/mV Meter

with Sensor Check

Hanna Instruments reserves the right to

modify the design, construction and appear-

ance of its products without advance notice.

Advertising