SALINTEST

HI 98203

Salinity Concentration
Pocket-sized Meter

GENERAL INFORMATION

The SALINTEST utilizes a sodium ion-selective
glass electrode to determine the activity of sodium
ions in solution.
When dissolved, NaCl ionizes to form Na

+

and Cl

-

ions. Measurement of Na

+

ion concentration is

an indicator of the concentration of NaCl.
The relationship between the SALINTEST read-
ing and the g/L NaCl scale is shown in the
logaritmic chart below.
A double junction reference is used to ensure a
highly stable reading.
If you suspect that the calibration has drifted,
you can recalibrate the meter by using HI 7081
or a solution of known concentration (adjust the
reading with the calibration trimmer).

HOW TO USE THE CHART

• Locate the pNa reading on the horizontal axis

of the chart.

• Move vertically upwards to intersect with the

45º line.

• Move horizontally and read the “g/L of NaCl”

value.

E.g. Display = 0.42

g/L NaCl = 22