Hanna Instruments HI 98180 User Manual

Page 63

Advertising
background image

63

Follow the Storage Procedure above.

pH CLEANING PROCEDURE

General

Soak in Hanna

HI 7061 or HI 8061 General

Cleaning Solution for approximately ½ hour.

Protein

Soak in Hanna

HI 7073 or HI 8073 Protein

Cleaning Solution for 15 minutes.

Inorganic

Soak in Hanna

HI 7074 Inorganic Cleaning Solution

for 15 minutes.

Oil/grease

Rinse with Hanna

HI 7077 or HI 8077 Oil and Fat

Cleaning Solution.

IMPORTANT: After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water, refill the reference chamber with
fresh electrolyte (not necessary for gel-filled electrodes) and soak the electrode
in

HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking

measurements.