Hanna Instruments HI 98150N User Manual

Page 14

Advertising
background image

14

• If “

WRONG” tag and “OLd” message are displayed on

the secondary LCD, an inconsistency exists between new
and previous (old) calibration data. Clear old calibration
parameters and proceed with calibration from the current
calibration point. The instrument will keep all confirmed
values during current calibration.

• With one-point calibration there is no “

Condition” and only

the frame is shown. Calibration time out is active.

• Each time a buffer is confirmed, the new calibration

parameters replace the old calibration parameters of
the corresponding buffer.

If current confirmed buffer has no

correspondence in the existing stored
calibration and this is not full, the
current buffer is added to the existing
stored calibration.

If the existing stored calibration is

full (five calibration points), after confirming the last
calibration point, the instrument will ask which buffer will be
replaced by current buffer.

Press the

ARROW keys to select another buffer to be replaced.

Press

CFM to confirm the buffer that will be replaced.

Press

CAL to exit this function. In this case, the buffer

calibration will not be stored.

Note: The replaced buffer is not removed from the calibration list and

can be selected for the next calibration.

WORKING WITH CUSTOM BUFFERS
If a custom buffer was configured in SETUP menu, it can be selected for
calibration by pressing the

ARROW keys. The “CUSTOM” tag will be

displayed blinking on the LCD.
Press

2nd then Setup if you want to adjust

the buffer value to it’s value at the current
temperature. The buffer value, displayed on
the secondary LCD, will start blinking.
Use the

ARROW keys to change the buffer value.

After about 5 seconds, the last buffer change
will be updated. Press

2nd then Setup if you want to change it again.

Notes: • Custom buffer value can be adjusted in a ±1.00 pH window,

around the set value.

• If you want to return to the set custom buffer value, simply

press the

ARROW keys ( then ) to select again the

custom buffer.