Errors and warnings – Hanna Instruments HI 96101 User Manual

Page 12

Advertising
background image

12

Under Range:

A blinking “0.00” indicates that the sample

absorbs less light than the zero reference. Check the procedure
and make sure you use the same cuvette for reference (zero) and
measurement.

Inverted cuvettes:

The sample and the zero cuvette are inverted.

Zero:

A zero reading was not taken. Follow the instructions of

the measurement procedure for zeroing the meter.

b) on sample reading

ERRORS AND WARNINGS

The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.

a) on zero reading

Light High:

There is too much light to perform a measurement.

Please check the preparation of the zero cuvette.

Light Low:

There is not enough light to perform a measurement.

Please check the preparation of the zero cuvette.

No Light:

The instrument cannot adjust the light level. Please

check that the sample does not contain any debris.

Advertising