Troubleshooting guide, Symptoms problem solution – Hanna Instruments HI 9321 User Manual

Page 18

Advertising
background image

1 8

TROUBLESHOOTING GUIDE

Symptoms

Problem

Solution

The meter is slow in
responding or gives
faulty readings

The meter does not
accept the 2

nd

buffer

solution for calibration

The reading drifts

Display shows:

Display shows:

Display shows:

Display shows:
"WRONG

BUF

1

"

and

"WRONG "

Display shows:

"WRONG

BUF

2

"

and

"WRONG "

The meter does not
work with the tem-
perature probe

The meter fails to cali-
brate or gives faulty
readings

The electrode is not
working or the reference
junction is clogged

Out of order pH elec-
trode

Defective pH electrode

Out of range pH scale

Out of range tempera-
ture scale

Out of range mV scale

Erroneous buffer solution
used for offset calibra-
tion

Defective electrode

Erroneous buffer solution
used for slope calibra-
tion

Defective electrode

Out of order temperature
probe

Out of order pH elec-
trode

Leave the electrode in a stor-
age solution after cleaning
the junction. If problem per-
sists, replace the electrode

Follow the cleaning proce-
dure. If this doesn't work
replace the electrode

Replace the electrode

a) Recalibrate
b) Make sure the pH

sample is in the 0 to 14
range

c) Check the electrolyte level

and the general state of
the pH electrode

Make sure the temperature
is in the 0 to 100°C range
and the temperature probe
is plugged in

Electrode not connected

Make sure the buffer setting
is correct and the solution is
fresh. Replace the buffer if
necessary

Replace the electrode

Make sure the buffer setting
is correct and the solution is
fresh. Replace the buffer if
necessary

Replace the electrode

Replace the probe

Replace the electrode

Advertising